Product Description
The Edwards Active Digital Controller (ADC) is a compact single gauge controller and display. It features a bright LED display and simple push-button controls. The ADC automatically recognizes compatible Edwards gauges, loads the appropriate look-up table and displays the pressure in commonly used vacuum units.
The enhanced controller supports two gauges - it has two variable hysteresis set-points which are linked to 48 V d.c. 1 A changeover relays and two 0-10 V d.c. analog outputs. To aid system integration, the enhanced controller is provided with an RS232 interface. When combined with a suitable gauge, such as the Edwards APGXH Convection Pirani or Wide Range Gauge (WRG), the ADC represents a cost effective means of monitoring and controlling process vacuum in a broad range of applications.
Features and Benefits
- Plug and measure operation means you simply plug in the mains supply, connect the gauge and ADC displays the measured pressure
- The ADC supports Edwards gauges with a total measuring range of 2000 to 1 x 10-9 mbar (1500 to 7.5 x 10-10 Torr)
- ADC and Edwards gauges address a broad spectrum of laboratory and industrial applications
- Bright LED display gives clear, long distance readability
- The ADC displays units in mbar, Torr, Pascal or Volts
- Small 1/8 DIN enclosure, may be panel or bench mounted
- Simple push-button control makes operation highly intuitive
Technical Data
Active gauge compatibility | APG100, APG-L, APG-MP, APG-M, APGX-H, APGX-L, AIM-S, AIM-X & ASG. |
Display | High brightness green LED display (0.47 inch high) |
Units - mbar/Torr/Pa/Volts | |
Electrical supply | 100 to 240V AC 47 to 63 Hz |
Serial interface | RS232 serial interface allows the ADC to be linked to a PC, PLC or OEM system controller. RS232 Protocol - 9600 baud, 1 stop bit, 8 data bits, no parity. |
Logic Interface | May be used either to link to system relays, or a higher-level control system and includes: |
Analog outputs | One 0-10V DC output signal per gauge channel |
Set-points | One user programmable, variable hysteresis set point per gauge channel. |
Control Relays | One volt-free, single pole change-over relay per gauge channel, tied to the respective programmable set-point: 1A @ 48V DC 2A @ 24V DC |
Dimensional Data | |
Panel cut-out | 92+0.8 x 45+0.6 mm (3.62inch x 1.77 inch) to DIN43700 |
Panel thickness | 1.5mm (0.06 inch) Min |
Weight | 0.33Kg |
Operating and storage data | |
Operating temperature | +0 to +40°C |
Storage temperature | -30 to +70°C |
Max ambient operating humidity | 90% RH non condensing at 40°C |
Standards | |
Electrical safety | BS EN 61010-1 |
Electrical noise immunity | BS EN 61326 (Industrial location, class B emissions) |
Flame retardant case materials | UL94 V1 |