Product Description
Overview
Features and Benefits
- Compatible with TIC turbo and instrument controller
- Multiple communication modes available including parallel mode, which allows traditional control signal interface and serial mode, which allows RS232 interface
- Numerous user configurable parameters via RS232 interface including vent valve control settings, normal speed indication, stand-by running speed, programmable power settings, power consumption and pump temperature.
- Enhanced monitoring capability including pump speed, power and temperature.
Technical Data
Inlet flange | DN63ISO-K, DN63CF, DN40NW or DN100ISO-K |
Pumping Speed | |
N2 ISO63/63CF (DN40NW)(ISO100) | 61 ls-1 (42 ls-1) (66 ls-1) |
He ISO63/63CF (DN40NW)(ISO100) | 57 ls-1 (49 ls-1) (59 ls-1) |
H2 ISO63/63CF (DN40NW)(ISO100) | 53 ls-1 (48 ls-1) (54 ls-1) |
Compression ratio | |
N2 | >1 × 1011 |
He | 1 x 106 |
H2 | 5 x 104 |
Ultimate Pressure (mbar) | |
With RV backing pump ISO/CF | -9 / 5 x 10-10 |
With diaphragm backing pump ISO/CF | -8 / 5 x 10-9 |
Outlet flange | DN16NW |
Recommended backing pump* | RV5/nXDS6i |
Vent port | 1/8 inch BSP |
Purge port | 1/8 inch BSP |
Max continuous inlet pressure (light gas pumping)† | |
Water cooling (water at 15 °C, ambient temp at 40 °C) | 2 × 10-2 mbar |
Forced air cooled, 35 °C ambient | 1 × 10-2 mbar |
Pump rotational speed | |
Nominal rotational speed | 90000 rpm |
Standby rotational speed | Variable from 49500 to 90000 rpm (63000 rpm default) |
Programmable power limit settings | Variable from 50-120W (80W default) |
Start time to 90% speed | 110 s ‡ |
Analogue outputs | Rotational speed; Power consumption; Pump temp; Controller temp |
Cooling method# | Forced air / water |
Ambient air temperature for forced air cooling | 5 - 35 °C |
Min cooling water flow rate (water 15 °C) | 15 l h-1 |
Water temp range | 10 - 20 °C |
Max inlet flange temp | 100 °C |
Operating attitude | Vertical and upright, through to Horizontal |
Noise level at 1 metre | |
Max magnetic field pump can tolerate | 5 mT |
Recommended controller | TIC100 turbo and instrument controller |
Quiescent electrical power | 10 W |
* A smaller backing pump may be used depending on application. A suitable diaphragm pump with ultimate † Above this inlet pressure, rotational speed drops to below nominal.
‡ Power limit set to 80 W.
# Air and water cooling accessories must be ordered separately
Technology
Quiescent electrical power is the nominal power dissipated by a pump operating normally at full rotational speed and with low gas throughput (inlet pressure below the 10–3 mbar range). During the run-up time, or when operating at high gas throughput or above the critical backing pressure, the pump power dissipation will rise and approach the maximum power output for the EXC Controller used. Critical backing pressure for conventional turbomolecular pumps is approximately 0.1 to 0.2 mbar.
Compression ratio is determined by the rotational speed, the number of pump stages and the molecular weight of the pumped gas. It is higher for heavier gases which explains why the suppression of hydrocarbon backstreaming is so effective and why the ratio for hydrogen is important for ultra high vacuum applications.
Ultimate pressure measured according to Pneurop standards, is the lowest pressure achieved in the test system, 48 hours after bakeout. The system is backed only by a two-stage rotary vane pump. Fluoroelastomer inlet seals are used with ISO-flanged pumps and metal seals are used with CF-flanged pump models.
Bearing and suspension technologies
We use two basic technologies: magnetic bearings and mechanical ceramic ball bearings. Ceramic bearings, which are lubricated for life by either grease or oil, have replaced conventional steel bearings. The silicon nitride ceramic balls are lighter, harder and smoother than steel equivalents, leading to longer life and lower vibration characteristics. Reliability is increased because the ball and race materials are different, which prevents micro pitting.
Magnetic bearings further increase reliability. Our EXT turbomolecular pumps up to 540 l s-1 use a hybrid bearing arrangement with a permanent magnet upper bearing and an oil or grease lubricated ceramic lower bearing. The EXT range also includes the EXT250M where the rotor is entirely supported by magnetic bearings. This offers additional advantages:
• Oil free - No hydrocarbon contamination from the turbomolecular pump
• Low maintenance - No bearing contact eliminates mechanical wear
• Low vibration - Typically an order of magnitude lower than conventional
bearings
• Reduced cooling - Ambient cooling is sufficient for many applications
• Any orientation - Pump can be mounted in any position
Rotor technologies
We use three basic technologies:
• conventional full stack turbomolecular (typically 12 stages)
• conventional short stack turbomolecular (typically 8 stages)
• compound molecular (combining turbomolecular and drag stages)
In addition, EXT pumps up to 540 l s-1 use monobloc rotors machined from solid bar by computer controlled high speed milling machines. This technology produces stable, rigid rotors and allows virtually unlimited design flexibility for optimum vacuum performance.
Motor technology
EXT pumps use brushless DC motors and are available in 24 and 80 volt variants. For the 24 volt pumps the TIC line of controllers are available with the added benefit of integrated instrument controllers. For the 80 volt pumps you can choose from our EXC line of controllers to optimize the performance and cost options for your application. The Controllers incorporate a regenerative back-up supply which provides power in the event of electrical supply failure to keep the vent-valve closed for several minutes.
Corrosive applications
For maximum life and reliability in the exacting process conditions encountered in semiconductor wafer processing applications, we recommend that you use turbomolecular pumps from our Edwards STP-C and STPHC series. These Maglev pumps have magnetic bearings and are ideal for these harsh duty applications.
Purge port
The EXT pumps (with the exception of the EXT70) all have purge-ports which can be used to purge the motor and bearing cavity with an inert gas (such as nitrogen). We recommend that you purge the pump when you pump corrosive and abrasive gas mixtures or those with an oxygen content over 25. You can use our PRX10 purge-restrictor to set the purge gas flow rate. This typically adds up to 25 sccm to the total gas load and the backing pump must be sized accordingly.
Venting
To maintain the cleanliness of your vacuum system, we recommend that you vent a turbomolecular pump at or above half rotational speed, when the rotor is still spinning fast enough to suppress any backstreaming of hydrocarbons from the backing line. The vent port on the EXT pump is part way up the rotor stack to ensure maximum cleanliness even with fluoroelastomer sealed vent-valves. Each pump is supplied with a manual vent-valve. If you use this manual valve care must be taken not to open it too quickly, especially if the system volume is small (typically less than the approximate volume of the turbomolecular pump), because if the rate of pressure rise is too high, the pump bearing life may be reduced.
In a small volume system, the rate of pressure rise will be greater than in a large volume for a given vent flow rate, and it may be necessary to restrict the vent gas flow. We offer the VRX range of vent restrictors which you can fit to your EXT pump.
Since the rate of pressure rise cannot be accurately controlled by the manual vent-valve, we recommend that, unless you fit a suitable VRX restrictor to the vent port, you must wait until the turbomolecular pump has slowed down to 50 speed, as indicated by the controller, before you open the manual vent-valve.
The maximum rate of pressure rise varies by pump model, and the Instruction Manual supplied with an EXT pump gives further guidance on this, and the size of vent restrictor needed to meet the fastest pressure rise allowed.
Control of the rate of venting is particularly important with pumps using fully magnetic bearings, otherwise the safety bearings may be damaged. The manual vent-valve can be replaced with a TAV solenoid valve driven by the EXC Controller to allow venting after a 2 second delay on shut-off, or delaying vent until the rotational speed has dropped to 50. The EXC Controller can also control the TAV vent-valve in the event of power or pump failure. You can choose from two solenoid vent-valve options; the TAV5 which covers most auto-venting applications, and the TAV6 which has a higher conductance than the TAV5 and is designed either for use on larger chambers (typically with a volume greater than 10 liters), or when you want to use a two-stage venting procedure for the fastest possible vent times. For two-stage venting you need two TAV valves. By using the appropriately restricted flow for the first stage vent-valve you can start venting when the EXT pump is still at full rotational speed. Once the pump has slowed to half rotational speed you can then introduce higher flow rates from the second stage vent-valve.
Inlet-screen
An inlet-screen is fitted as standard to all EXT pumps. The inlet-screen prevents debris from falling into the pump-inlet. In addition, the inlet screen prevents you from coming into contact with the blades of the pump when it is disconnected from your vacuum system.
Cooling
For most applications, we recommend that you use forced-air cooling with the appropriate ACX air-cooler connected to your EXT pump. If the ambient temperature is below 30 °C, the EXT70, EXT70H and EXT250M pumps can be cooled by free convection (dependent on gas load and backing pressure). NB: high gas load, high backing pressure and rapid cycling require more cooling.
Water cooling reduces the running temperature of the pump motor and bearings and is particularly recommended when you operate the EXT with a continuous high throughput (that is, inlet pressure above 1 x 10-3 mbar)
or when you bake the EXT pump to above 70 °C (measured at the inlet flange).
Scope of supply
For end users desiring front panel controls and indications when using an 80 volt pump a minimum operating system requires you to order an EXT pump, an EXC controller and a pump to controller cable. For a 24 volt pump a minimum operating system requires you to order an EXT pump, a TIC controller and an EXDC controller. (The EXDC controller is not required when using an EXT75DX or EXT255DX). Each EXT pump is supplied with an inlet-screen, elastomer inlet seal or copper gasket (as appropriate), manual vent valve and water cooler. Each EXC/TIC controller is supplied with a 2 m unterminated electrical supply cable.
Technical Data
Inlet flange | DN160ISO-K, DN160CF |
Outlet flange | DN25NW |
Pumping speed | |
N2 | 400 ls-1 |
He | 390 ls-1 |
H2 | 325 ls-1 |
Compression ratio | |
N2 | >1 × 1011 |
He | 1 x 108 |
H2 | 5 x 105 |
Ultimate pressure with RV backing pump ISO/CF | -8 /-10 mbar |
Recommended backing pump* | RV12/nXDS10i |
Vent port | 1/8 inch BSP female |
Purge port | 1/8 inch BSP female |
Max continuous inlet flow (Nitrogen)† | |
Water cooling (water at 15°C, ambient temp at 40°C) | 105 sccm |
Forced air cooled, 35 °C ambient | 90 sccm |
Pump rotational speed | |
Nominal rotational speed | 60000 rpm |
Standby rotational speed | Variable from 33000 to 60000 rpm (42000 rpm default) |
Programmable power limit settings | Variable from 50-200W (160W default) |
Start time to 90% speed | 180s‡ |
Analogue outputs | Pump rotational speed; Power consumption; Pump temp; Controller temp |
Cooling method# | Forced air / water |
Ambient air temp for forced air cooling | 5 - 35°C |
Min cooling water flow rate (water 15°C) | 15 l h-1 |
Water temp range | 10 - 20°C |
Max inlet flange temp | 70°C |
Operating attitude | Vertical and upright, through to Horizontal +/- 2° |
Noise level at 1 metre | |
Max magnetic field pump can tolerate | 5 mT |
Recommended controller | TIC200 turbo and instrument controller |
Quiescent electrical power | Typically 20 W |
* A smaller backing pump may be used depending on application. A suitable diaphragm pump with ultimate † Above this inlet pressure, rotational speed drops to below nominal.
‡ Power limit set to 160 W.
# Air and water cooling accessories must be ordered separately